updatedAt: 2026-05-16T00:03:40.000Z
agentTools:
  projectIndex: https://developers.tetrascience.com/llms.txt
---

# Tetra SDC Connector

[llms-txt]: # "For AI agents: the complete documentation index is at https://developers.tetrascience.com/llms.txt — markdown versions of any page are available by appending .md to the URL."

The Tetra SDC Connector is an improved version of the [Legacy Tetra SDC Connector](https://developers.tetrascience.com/docs/agu-sdc) . The Connector is a standalone, containerized application that enables **bi-directional** communication between the Tetra Data Platform (TDP) and AGU Sm\@rtLine Data Cockpit (SDC). SDC is a middleware that captures measurements from sensors and analyzers and exports them through REST API export interfaces. Users can configure the Connector to periodically poll export interfaces for sample and measurement results. The Connector also allows integrating systems to submit measurement order requests to SDC through the TDP [Command Service](https://developers.tetrascience.com/docs/command-service).

# Prerequisites

* AGU SDC version 3.0 or higher
* **(Optional)** [Tetra Hub](https://developers.tetrascience.com/docs/tetra-hub-v2). You can deploy the Connector to an on-premises Tetra Hub instead of the cloud if networking access between SDC and the TDP is restricted. When deployed to a Tetra Hub, internal networking/firewall rules to allow communication between the Tetra Hub and SDC must be configured. For more information, contact your customer success manager (CSM).

# Design Overview

AGU SDC exports measurements and sample results through REST API export interfaces. The export interfaces can be created by SDC admins and configured to export results from selected instruments in SDC. These export interfaces behave in a queue-like fashion.

The Tetra SDC Connector v2 periodically polls these interfaces for available results, and then uploads them to the TDP. Data ingestion capabilities were expanded from the Tetra SDC Connector v1 to support the upload of both measurement and sample results. Measurement orders can also be submitted to SDC programmatically by issuing commands through the TDP [Command Service](https://developers.tetrascience.com/docs/command-service).

## Architecture

The following diagram illustrates an example Tetra SDC Connector v2 deployment to replatform data from SDC to a single TDP organization:

<Image align="center" alt="SDC Connector V2 Architecture" border={false} caption="SDC Connector V2 Architecture" title="1_sdc.png" src="https://files.readme.io/ca72566-sdc_conn_architecture.png" width="90% " />

> 📘 Working with Multiple TDP Organizations
>
> When scaling an SDC integration across multiple TDP organizations, it's recommended that you create a separate export interface and Connector for each organization. Using different Connectors and export interfaces for each organization allows the TDP to handle the queue-like nature of SDC export interfaces with high data integrity. Connectors are also coupled to their TDP organizations.

### Dataflow Example

The following diagram shows an example Tetra SDC Connector v2 dataflow:

<Image align="center" alt="SDC Connector V2 Architecture" border={false} caption="SDC Connector V2 Dataflow" title="1_sdc.png" src="https://files.readme.io/99c4bb4-sdc_conn_dataflow.png" width="90% " />

# Operational Guide

To set up an integration between SDC and TDP using the Tetra SDC Connector v2, do the following.

## Configure the SDC Export Interface

> 🚧 IMPORTANT
>
> It's recommended that you create and configure a dedicated export interface in SDC for each Tetra SDC Connector v2 used in your TDP deployment. To mirror your SDC data across your TDP organizations, you must deploy a separate Connector for each of your organizations.

To create and configure export interface(s) in AGU SDC, do the following:

1. Sign in to AGU SDC as an admin. Open the **Export Interfaces** page by doing the following:
   * In the left navigation menu, choose **Settings**.
   * Choose **Interfaces**.
   * Select the **Export Interfaces** icon.

<Image align="center" border={false} width="90% " src="https://files.readme.io/0ed7bcd-agu_sdcc_config_export_interface.png" />

2. Select the **New** button. The Create export interface workflow appears.

<Image align="center" border={false} width="90% " src="https://files.readme.io/7d56a19-agu_sdc_create_export_interface_2.png" />

3. Define the export interface by doing the following:
   * For **Description**, enter a name for the export interface (for example, **Exports-to-TDP**)
   * For **Export type**, select **RestServer**.
   * For Event type, choose whether the interface exports measurement results (**Measurement**) or sample results (**SampleResult**).
   * Choose **Save**. The Export Interface Exports-to- workflow appears.
4. Edit your export interface to match the following **Common** configurations:

<Image align="center" border={false} width="90% " src="https://files.readme.io/673fe32-agu_sdc_create_export_interface_3.png" />

* For **Active**, select the check box.

> 📘 NOTE
>
> Active interfaces add data to the queue. Inactive interfaces do not add data to the queue. However, inactivating a queue doesn't delete results from the queue; it only prevents new results from being added.

* For **All parameters**, select the check box.

> 📘 NOTE
>
> Selecting **All parameters** is required for proper file versioning in the TDP.  If the **All parameters** check box isn't selected , then the exports will create partial files in the TDP, where each file contains a fraction of all results.

* **(Optional)** For **Diagnostics**, select the check box. This option isn't required, but is helpful for troubleshooting purposes.

5. Configure your export interface to export data from specific divisions in the **Division assignment** configurations.

<Image align="center" border={false} width="90% " src="https://files.readme.io/08f3379-agu_sdc_create_export_interface_5_division.png" />

6. Configure your export interface to export data from a specific instrument make and model in the **Device type assignment** configurations.

> 🚧 IMPORTANT
>
> AGU SDC can't export data from a specific instrument ID. All instruments of a specific make and model export together.

<Image align="center" border={false} width="90% " src="https://files.readme.io/8251780-agu_sdc_create_export_interface_6.png" />

## Create a Tetra SDC Connector v2

To create a Tetra SDC Connector v2, do the following:

1. [Sign in to the TDP](https://developers.tetrascience.com/docs/sign-in-and-out-of-tdp-and-manage-password). Then, in the left navigation menu, choose **Data Sources** and select **Connectors**. The Connectors page appears.
2. Follow the instructions in [Create a Pluggable Connector](https://developers.tetrascience.com/docs/create-configure-and-update-connectors#create-a-pluggable-connector). For CONNECTOR TYPE, make sure that you select **common AGU SDC**.

> 📘 NOTE
>
> After the Connector is created, the initial **MODE** status is set to **IDLE** by default. To activate the Connector, you must configure its settings. For instructions, see the **Configure the Connector** section.

## Configure the Connector

> 🚧 IMPORTANT
>
> It's recommended to have a 1:1 relationship between the Tetra SDC Connector v2 in the TDP and your SDC export interfaces. If two Connectors poll the same export interface, then the integration may experience unpredictable behavior in SDC and the TDP when independent Connectors simultaneously poll the same export interface queue. Make sure that you review your Connector configurations carefully.

On the **Connectors** page, select the name of the connector that you created. Then, select the **Configuration** tab to configure the required settings.

### Configuration Parameters

The following tables lists the configuration parameters that are required for the Tetra SDC Connector v2.

| Configuration Parameter                     | Description                                                                                                                                                                        |
| :------------------------------------------ | :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
| **SDC URL**                                 | URL for the SDC server including protocol, hostname and port. If port is not specified, the default port for protocol will be used (for example, <https://yoursdcserver.com:8080>) |
| **Username**                                | Username of the SDC user with permissions to create measurement orders and request data from an export interface.                                                                  |
| **Password**                                | Password for the user.                                                                                                                                                             |
| **Client ID**                               | If provided, the Connector will use it for OAuth 2.0 authentication                                                                                                                |
| **Client Secret**                           | If provided,  the Connector will use it for OAuth 2.0 authentication                                                                                                               |
| **Measurement Result Export Interface IDs** | Comma-separated list of SDC Measurement Result Export Interface IDs to poll (for example, **2, 4**)                                                                                |
| **Sample Result Export Interface IDs**      | Comma-separated list of SDC Sample Result Export Interface IDs to poll (for example, **1, 3**)                                                                                     |
| **Polling Interval**                        | Interval, in minutes, of how often data will be uploaded to the TDP. The minimum value is **1** minute. The default value is **10** minutes.                                       |
| **Max Results per Interval**                | Maximum number of results to upload per polling interval. If not specified, no limit is enforced.                                                                                  |
| **Disable SSL Certificate Verification**    | If checked, SSL/TLS certificate verification is deactivated. The default value is **false**.                                                                                       |

## Review and Edit the Connector's Information

The **Information** tab on the Connector Details page displays information about the Connector, the files pending, successfully uploaded, or failed. It also provides functionality to edit a Connector’s information, metadata, and tags.

To edit the Connector's information, select the **Edit** button on the Connector Details page's **Information** tab.

For more information, see [Review and Edit a Pluggable Connector's Information](https://developers.tetrascience.com/docs/create-configure-and-update-connectors#review-and-edit-a-pluggable-connectors-information).

### Connector Details

The following table provides the Connector details available on the **Information** tab on the Connector Details page.

<Table align={["left","left"]}>
  <thead>
    <tr>
      <th>
        Field Name
      </th>

      <th>
        Description
      </th>
    </tr>
  </thead>

  <tbody>
    <tr>
      <td>
        **CONNECTION**
      </td>

      <td>
        * **ENABLED**
        * **DISABLED**
          * *Note:** When disabled, no container is running for the connector.
      </td>
    </tr>

    <tr>
      <td>
        **NAME**
      </td>

      <td>
        Name of the connector
      </td>
    </tr>

    <tr>
      <td>
        **NETWORK**
      </td>

      <td>
        * **ONLINE:**Container is created and can be reached by the TDP.
        * **OFFLINE:**Container cannot be reached by the TDP.
      </td>
    </tr>

    <tr>
      <td>
        **HEALTH**
      </td>

      <td>
        * **HEALTHY:**Everything is fine, connector doesn’t see any issues.
        * **WARNING:** There is some issue but it’s not critical.
        * **CRITICAL:** Critical issue that prevents connector from running. Connector cannot upload data or execute some commands even in case its operational status is running.
        * **N/A:** The connector doesn’t have a health status yet. This generally indicates that a connector was created but has not been configured.
      </td>
    </tr>

    <tr>
      <td>
        **MODE**
      </td>

      <td>
        * **IDLE:** Connector is enabled, but should not be uploading data. The connector may accept and execute commands. When the connector is first created, its **MODE** status is **IDLE** by default.
        * **RUNNING:**Connector is enabled, should be running and uploading data. The connector may accept and execute commands.
        * **DISABLED:** Connector container is shut down.
      </td>
    </tr>

    <tr>
      <td>
        **HOST**
      </td>

      <td>
        * **Cloud:**Connector was deployed as a standalone server in the Cloud
        * **Hub:**Connector was deployed in Tetra Hub
      </td>
    </tr>

    <tr>
      <td>
        **CONNECTOR ID**
      </td>

      <td>
        The connector’s ID
      </td>
    </tr>

    <tr>
      <td>
        **TYPE**
      </td>

      <td>
        Connector type
      </td>
    </tr>

    <tr>
      <td>
        **VERSION**
      </td>

      <td>
        The connector’s version
      </td>
    </tr>

    <tr>
      <td>
        **LAST SEEN**
      </td>

      <td>
        Last NETWORK check
      </td>
    </tr>

    <tr>
      <td>
        **DESCRIPTION**
      </td>

      <td>
        User-defined connector’s description
      </td>
    </tr>
  </tbody>
</Table>

#### Connector Information That You Can Edit

The following table provides the connector information that can be edited after a Connector is created.

| Field Name                    | Description                                                                                 |
| :---------------------------- | :------------------------------------------------------------------------------------------ |
| **Disabled/Enabled** (toggle) | Toggle this field to enable/disable the connector                                           |
| **NAME**                      | Connector’s name                                                                            |
| **DESCRIPTION**               | Connector’s description                                                                     |
| **CONNECTOR VERSION**         | Connector version                                                                           |
| **Add / Edit Attributes**     | Select this option to add or edit labels, metadata, and tags associated with the connector. |

## Metrics

After a Connector instance is created and running, you can monitor the Connector’s health by selecting the **Metrics** tab on the Connector Details page. The Metrics tab displays the Connector's container metrics as well as aggregated states on total files scanned and uploaded as well as files that returned errors or are pending.

# Integrate Data

Measurement and sample results that are exported to configured exported interface(s) will be uploaded to TDP. After data is uploaded to the TDP, you can enrich and harmonize the data by using any of the following available scripts, protocols, and intermediate data schemas (IDS):

* Task script: `common/sdc-measurement-raw-to-ids:v2.0.0`
* Protocol: `common/sdc-measurement-raw-to-ids:v2.0.0`
* IDS: `common/agu-sdc:v2.0.0`

## Submit Measurement Order

Measurements can be created programmatically in AGU SDC through a TetraScience `CreateMeasurementOrder` command.

To create a TetraScience `TetraScience.SDCConnector.CreateMeasurementOrder` command, an API call must be made to the [Create Commands API](https://developers.tetrascience.com/reference/ag-create-command). The Connector passes the `payload.measurementOrderRequestBody` to the SDC REST API.

### Required parameters

* `SampleNumber`
* `SampleType`
* `DeviceNumber`

The following command data can be posted to the Command API to make a measurement in the SDC Sandbox:

```
{
    "targetId": "7e71ceea-0503-4031-9b56-97c686245e25",
    "action": "TetraScience.SDCConnector.CreateMeasurementOrder",
    "expiresAt": "2025-12-30T23:58:43.749Z",
    "payload": {
        "ttl": 600,
        "measurementOrderRequestBody": {
            "SampleNumber": "Sample987",
            "ExternalId": "LIMS12345",
            "SampleType": "Advanced Instruments OsmoTECH PRO default",
            "DeviceNumber": "AIOsmotechPro123456"
        }
    }
}
```

| Parameter Name                        | Required | Description                                                                                                                                                                                                                                             |
| :------------------------------------ | :------- | :------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| `targetId`                            | Yes      | Connector UUID from the Connector Details page                                                                                                                                                                                                          |
| `action`                              | Yes      | Must be `TetraScience.SDCConnector.CreateMeasurementOrder`                                                                                                                                                                                              |
| `expiresAt`                           | Yes      | Deadline after which incomplete commands are considered expired                                                                                                                                                                                         |
| `payload.ttl`                         | No       | Numeric timeout for the HTTP request, in seconds (  Connector default is 300,000)                                                                                                                                                                       |
| `payload.measurementOrderRequestBody` | Yes      | These contents are posted to SDC’s REST API `/SDC/Measurements`, and are handled by SDC directly.  The requestBody contents can vary based on which instrument is used. For more information, see the SDC User Manual and REST Configuration procedure. |

# Documentation Feedback

Do you have questions about our documentation or suggestions for how we can improve it? [Start a discussion in TetraConnect Hub](https://tetrascience.zendesk.com/hc/en-us/community/topics/38259263466893-Documentation-Feedback). For access, see [Access the TetraConnect Hub](https://developers.tetrascience.com/docs/access-the-tetrascience-knowledge-base).

> 📘 NOTE
>
> Feedback isn't part of the official TetraScience product documentation. TetraScience doesn't warrant or make any guarantees about the feedback provided, including its accuracy, relevance, or reliability. All feedback is subject to the terms set forth in the [TetraConnect Hub Community Guidelines](https://tetrascience.zendesk.com/hc/en-us/articles/26916126183309-TetraConnect-Hub-Community-Guidelines).