Sotax DisiTest50 (DT50)

📘

NOTICE

DT50 can only be configured to send data to one destination. If you connect the DT50 to the IoT box, it can’t send data to the printer.

Integration method: AnyLink IoT Agent (IE Pro model)
Cables: 1x Ethernet cable

Installation

  1. Connect an AnyLink box’s LAN port to DT50, or to the network switch the DT50 is using (if there is one).
  2. The default AnyLink IP address is 192.168.101.204 with subnet mask 255.255.255.0. On the DT50, change the DT50 IP address to be in the same subdomain (for example, you can change the last three digits in the default address to any number from 3-254, but can't use the default 204: 192.168.101.xxx). If you have an auxiliary tower, make sure that you are not using the same number as the auxiliary tower either.
  3. Change the printer setting to label.
  4. Open an experiment and select Print to send the data to the IoT Agent device. If everything works correctly, you'll see a file appear on the Tetra Data Platform (TDP).